IntelliSpace Critical Care And Anesthesia

Health product recall

IntelliSpace Critical Care And Anesthesia

 

Brand(s)

Philips Medizin Systeme Boblingen Gmbh

Last updated

2026-09-16

Summary

Product

IntelliSpace Critical Care And Anesthesia

Issue

Medical devices – Performance

What to do

Contact the manufacturer if you require additional information.

Audience

Healthcare

Affected products

Affected products Lot or serial number Model or catalogue number
IntelliSpace Critical Care And Anesthesia Not applicable. VERSION J.06

Issue

In ICCA releases J.06.01 and J.06.02, calculations triggered by infusion-pump auto-charting may result in incorrect Intake Totals, Output Totals, Net Body Balance, Length of Stay (LOS) Balance, and related fluid-balance values.
This issue applies only to customers that have configured and use Auto Charting for Infusion Data for a Clinical Unit, with one or both of the following settings enabled:
• Automatic periodic charting of infusion rates and volumes.
• Automatic aperiodic charting of infusion rates and bolus volumes when a change occurs.
Customers that do not have either Auto Charting for Infusion Data setting configured and in use are not impacted by this issue.

Additional information

Details

Original published date: 2026-09-16

Alert / recall type

Health product recall

Category

Health products – Medical devices – Anaesthesiology

Companies

Philips Medizin Systeme Boblingen GmbH

Hewlett-Packard Strasse 2, Boblingen, Germany, 71034

Published by

Health Canada

Audience

Healthcare

Recall class

Type II

Recall date

2026-09-04

Identification number