Health product recall
IntelliSpace Critical Care And Anesthesia
Brand(s)
Philips Medizin Systeme Boblingen Gmbh
Last updated
2026-09-16
Summary
Product
IntelliSpace Critical Care And Anesthesia
Issue
Medical devices – Performance
What to do
Contact the manufacturer if you require additional information.
Audience
Healthcare
Affected products
| Affected products | Lot or serial number | Model or catalogue number |
|---|---|---|
| IntelliSpace Critical Care And Anesthesia | Not applicable. | VERSION J.06 |
Issue
In ICCA releases J.06.01 and J.06.02, calculations triggered by infusion-pump auto-charting may result in incorrect Intake Totals, Output Totals, Net Body Balance, Length of Stay (LOS) Balance, and related fluid-balance values.
This issue applies only to customers that have configured and use Auto Charting for Infusion Data for a Clinical Unit, with one or both of the following settings enabled:
• Automatic periodic charting of infusion rates and volumes.
• Automatic aperiodic charting of infusion rates and bolus volumes when a change occurs.
Customers that do not have either Auto Charting for Infusion Data setting configured and in use are not impacted by this issue.
Additional information
Details
Original published date: 2026-09-16
Alert / recall type
Health product recall
Category
Health products – Medical devices – Anaesthesiology
Companies
Philips Medizin Systeme Boblingen GmbH
Hewlett-Packard Strasse 2, Boblingen, Germany, 71034
Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Recall date
2026-09-04
Identification number
RA-82635
Get notified
Receive emails about new and updated recall and safety alerts.
Report a health or safety concern
- Date modified:
- 2026-09-16
Source: Originally published by Health Canada / Government of Canada.
Category: Health Products
Original alert: https://recalls-rappels.canada.ca/en/alert-recall/intellispace-critical-care-and-anesthesia
Reproduced under the Open Government Licence – Canada.